    Product Introduction of the thermal shock tester

    Application:

    This thermal shck test equipment is prohibited to flammable, explosive, easy to volatile substances sample test or storage, corrosive substance sample test or storage, biological sample test or storage, strong electromagnetic emission source sample test or storage.



    Test Space Option of the thermal shock tester

    Volume42L80L150L225L408L

    Inner Dimension

    (WxHxD)cm

    36x35x4050x40x4060x50x5050x75x6060x85x80

    Outer Dimension

    (WxHxD)cm

    138x180x155150x190x160190x200x195210x220x205225x250x220

    Key Specifications of the thermal shock tester

    Temperature range of test area-40℃~150℃, -60℃~150℃, -65℃~150℃ ( option)
    High temperature range: +60℃~150℃
    Low temperature range: -10℃~150℃
    Testing environment condition

    Environment temperature: +28℃

    Relative humidity: ≤ 85%

    In the condition of without sample in the sample chamber

    Preheating temperature range+60℃~+150℃
    Heating up time

    +60℃→+150℃

    ≤ 40min

    Note: The heating up time is the performance when

    the high temperature chamber individually operates.

    Low temperature chamber

    Note: The heating up time is the performance when

    the high temperature chamber individually operates.

    Precooling temperature range-55℃~-10℃
    Cooling time

    +20℃→ -55℃≤ 70min

    Note: The cooling time is the performance when

    the low temperature chamber individually operates.

    Testing methodSwitch pneumatic ventilation door 2 or 3 temperature zones.
    Temperature shock range-40℃~+150℃
    Temperature fluctuation±0.5℃
    Temperature departure±2.0℃
    Temperature recovery time≤ 5min

    Recovery condition

    (Chart One)


    Sample: plastic packaged integrated circuit(uniformly distribute)

    Sensor: the upper side of the sample

    High temperature exposure: +150℃ 30 minutes

    Environment temperature exposure: ---

    Low temperature exposure: -40℃ 30 minutes

    Sample weight: 2.5kg


    Structure of the thermal shock tester

    1. Heat preservation envelop enclosure:

    • Outer wall material: cold-roll steel sheet stoving varnish.
    • Inwall material: stainless steel plate SUS304.
    • Body heat preservation material: rigid polyurethane foam and glass fiber.
    • Door heat preservation material: glass fiber.

    2. Air conditioning channel: Fan, heater, evaporator, air door and temperature sensor.
    3. The standard configuration of chamber body: a lead drilling hole, locate on the left surface of the chamber body, φ50mm round hole.


    4. load capacity of test chamber basket (uniform distribution),

    2 sample basket, 5kg load capacity each(uniform distribution).


    5. Adjusting foot and adjusting wheel: 4 adjusting feet to support the test chamber, 4 adjusting wheels to move the test chamber.
    6. Thermal shock test chamber door: single open, left hinge, right handle (reference of Test chamber outline drawing).
    7. Circulating fan:


    • High temperature/ring temperature exposure: 3φ,400W, centrifugal
    • Low temperature exposure: 3φ,400W, centrifugal

    8. Control panel: Controller display screen, failure indicator, hour meter, R232 interface and over temperature protection setting device, total power switch.


    9. Thermal shock test machine room: Refrigeration unit, drainage device, exhaust fan.

    10. Power distribution control cabinet: main power leakage circuit breaker, controller, power distribution board, cooling fan, high and low temperature circle back fan motor.


    11. Heater: Nickel chromium alloy strip heater,
    Heater control mode: non-contacting periodic pulse width, SSR (solid-state relay).
    12. Power line hole and drain hole: at the back of the thermal shock chamber.



    Quality Thermal Shock Test Chamber Manufacturers for sale